Blue Beads 6FF (Affinity chromatography)
Blue Beads 6FF is a chromatography medium for the isolation of proteins,such as albumin, interferon, and blood coagulation factors. Blue Beads 6FF also binds several enzymes including kinases, dehydrogenases, and most enzymes requiring denyl-contatining cofactors, for example NAD+.
Blue Beads 6FF consists of 90µm beads of highly cross-linked 6% agarose, is highly stable and expand the range of suitable operating conditions.
Table 1. Characteristics of Blue Beads 4FF
|
Item |
Description |
|
Matrix |
Highly cross-linked 6% agarose |
|
Ligand |
Cibacron Blue 3G |
|
Capacity (/ml medium) |
>18 mg bovine serum albumin |
|
Particle size (μm) |
45-165 |
|
Maxi pressure |
0.3 MPa, 3 bar |
|
pH stability |
4-12 |
|
Storage buffer |
0.1 M K3PO4, 20% ethanol, pH 8.0 |
|
Storage |
2°C - 8°C |
2.1 Buffer Preparation
Water and chemicals used for buffer preparation should be high purity. It is recommended filtering the buffers by passing them through a 0.22μm or 0.45μm filter before use.
Purification of human serum albumin as an example.
Binding /Wash Buffer: 50mM Na2HPO4, 50mM Na-Citrate, pH7.0
Elution Buffer: 50mM Na2HPO4, 50mM Na-Citrate, 1-2M NaCl, pH7.0
The Binding/Wash Buffer and Elution Buffer can be changed according to the properties of the samples. The principle is load at low salt and elute at high salt.
2.2 Sample Preparation
It is recommended to filter the sample solution by passing them through a 0.22μm or 0.45 μm filter before use.
Blue Beads 6FF are widely used in industrial purification. Therefore, it involves the packing of various medium and low pressure chromatographic columns. The following describes the packing method of Blue Beads 6FF column.
2). Close the column outlet leaving the net covered with packing buffer.
3). Resuspend the beads stored in its container by shaking (avoid stirring the sedimented medium). Pouring the slurry down a glass rod held against the column wall will minimize the introduction of air bubbles.
If using a packing reservoir, immediately fill the remainder of the column and reservoir with packing buffer. Mount the adapter or lid of the packing reservoir and connect the column to a pump. Avoid trapping air bubbles under the adapter or in the inlet tubing.
velocity of approximately 400 cm/h (10 cm bed height, 25°C, low viscosity buffer).If the recommended pressure or flow velocity can not be obtained, use the maximum flow velocity the pump can deliver. This should also give a reasonable well-packed bed. Do not exceed 75% of the packing flow velocity in subsequent chromatographic procedures.
5). When the bed has stabilized, close the bottom outlet and stop the pump.
If using a packing reservoir, disconnect the reservoir and fit the adapter to the column. If using the column, carefully place the top filter on top of the bed before fitting the adapter.
6). With the adapter inlet disconnected, push the adapter down, approximately 2 mm into the bed, allowing the packing solution to flush the adapter inlet.
7). Connect the pump, open the bottom outlet and continue packing. The bed will be further compressed at this point and a space will be formed between the bed surface and the adapter.
8). Close the bottom outlet. Disconnect the column inlet and lower the adapter approximately 2 mm into the bed. Connect the pump. The column is now ready to use.
2.4 Sample Purification
1). Fill the syringe or pump tubing with binding buffer. Remove the stopper and connect the column to the syringe (with the provided connector), or pump tubing, “drop to drop” to avoid introducing air into the column. Remove the snap-off end at the column outlet.
2). Wash the column with at least 5 column volumes of binding buffer.
3). Apply the sample, using a syringe fitted to the connector or by pumping it onto the column.
4). Wash with 10 to 15 column volumes of binding buffer or until no material appears in the effluent.
5). Elute with 5-10 column volumes of elution buffer. Other volumes may be required if the interaction is difficult to break.
2.5 Analysis
Identify the fractions containing the target protein. Use UV absorbance, SDS-PAGE, or western blot.
In general, Blue Beads 6FF is well suited for reuse a number of times. When precipitation and protein aggregation cause the loss of velocity and combined loads, you need to clean the medium as follows.
Remove the precipitated proteins or denatured substance
Wash the column with 3-4 column volumes 0.1M NaOH, followed by washing the column with 3-4 column volumes of 70% ethanol or 2M potassium thiocyanate. Alternatively, wash the column with 2 column volumes of 6M guanidine hydrochloride.
Finally wash the column with at least 5 column volumes of binding buffer.
Remove the strong hydrophobic binding proteins, lipoproteins and lipids
Wash the column with 3-4 column volumes 70% ethanol. Alternatively, wash the column with two column volumes of detergent in a basic or acidic solution. Wash the residual detergent by 5 column volumes of 70% ethanol.
Finally wash the column with at least 5 column volumes of binding buffer.
4. Related Products
|
Product |
Cat. No. |
Size |
|
Blue Beads 6FF |
SA027005 SA027025 SA027100 SA027500 SA02701L SA02710L |
5 ml 25 ml 100 ml 500 ml 1 L 10 L |